Germicide Test Strips WaterCheck RC
Bleach chlorine solution is commonly used as a cleansing agent to disinfect medical instruments or equipments, such as dialysis machines. After the disinfection, chlorine is then rinsed off with deionized water or RO water before the instruments can be reused. The WaterCheck RC chlorine test strip provides an easy way to monitor the chlorine concentrations during the process of disinfection. Product Features: 1. Fast and sensitive.
2. Very consistent
3. Easy to read colors
4. Residual and potency strip in one
5. Long shelf life, stable at room temperature
Simple Steps to Use The strip is primarily designed for use in hemodialysis disinfection control. However, it can also be used in different applications. For disinfection of dialysis machines, bicarbonate mixing tanks or water distribution system, the final rinse water should have none or less than 0.5 ppm of residual chlorine according to the AAMI standards. The simple strip test can verify the compliance of the water quality.
Interpretation of the test results depends on the nature of the testing. Using the one second dip-and-read test procedure, the strip can detect as low as 0.2 ppm of residual chlorine and will show a recognizable green color. Although the maximal allowable residual chlorine contaminant level in the rinse water is 0.5 ppm by AAMI standards, in practice, all rinse water in hemodialysis setting is tested negative with the strip. If tested positive, the rinsing should be continued. If measurement of a more accurate chlorine level is desired, a 5 second test procedure should be followed by holding the strip under a slow running test water for 5 seconds and comparing the strip color with the color chart immediately. The strip color will be intensified at the end of 5 seconds. The length of the contact time of the strip with fresh test sample determines the final intensity of the strip color. The longer the contact time, the darker the strip color will be. The actual strip color may vary depending on how the test is done and the lighting condition of the room. The residual color blocks on the label are standardized with the 15 second test procedure and is considered semiquantitative only.
The strip is the most commonly used in detecting chlorine in water. However, the strip can also be used for detecting chlorine in other media, such as dialysate or bicarbonate solutions. In the blood, plasma or serum, chlorine will reacts with blood components such as ascorbic acid or other antioxidants, also with serum proteins and reduce the detectable chlorine level. What is detected in such media is the remaining of the chlorine rather than the initial level of chlorine contamination. In solvents or oil, the presence of chlorine can also be detected by the strip. However, the results should only be interpreted as qualitative. How to Chose the Right Chlorine Test Strip for Your Specific Application Since the test is based on the oxidative election transfer of chlorine molecules, chlorine will reacts to a variety of indicator system with different chemical properties. Because of this, not all the chlorine tests are created equal even when the same chemical indicator is used. They all differ in detection sensitivity, detection range, chemical specificity and thermal stability. IBT offers three chlorine test strips for different applications. The following table compares the main features and the suggested applications of each strip.
There are many other test strips which will react with chlorine but lack the sensitivity required for hemodialysis or other critical quality water applications. For detailed explanation, refer to the IBT Bulletin #402 “The Most Commonly Misused Chlorine Test Strips”
The WaterCheck RC test strips are packaged 100 strips in a bottle and 6 bottles in a box.
Order by phone, fax or email. Contact our Customer Service for more information or technical support. Customer Service: R: ec103
|
|
||||||||||||||||||
|
© 2005
Integrated Biomedical Technology. All rights reserved. 2931 Moose Trail,
Elkhart, IN 46514 U.S.A. Phone: (574) 264-0025, Fax: (574) 264-2787 Customer Service: (800) 490-5500 |
||||||||||||||||||||||